FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3203249
·
Received July 2, 2013
Report
- Report Number
- 3004209178-2013-95301
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- June 22, 2013
- Report Date
- June 24, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARM DURING BASIC OCCLUSION TEST DUE TO LOOSE DRIVE SUPPORT DISK. SCRATCHED LCD WINDOW NOTED DURING VISUAL INSPECTION. MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DRIVE SUPPORT CAP WAS LOOSE. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302342 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |