FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3203222 · Received July 2, 2013

Report

Report Number
3004209178-2013-95317
Event Type
Injury
Date Received
July 2, 2013
Date of Event
March 1, 2013
Report Date
June 12, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO LOW BLOOD GLUCOSE. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. REVIEWED THE PROGRAMMING AND FOUND THAT THE DEVICE WAS OFF AN HOUR. ASSISTED THE CALLER TO RESET THE TIME. REVIEWED THE PROGRAMMING, AND THE RESERVOIR WAS SHOWING THE SAME AMOUNT OF INSULIN AS SHOWN ON THE STATUS SCREEN. PERFORMED THE HIGH PRESSURE TEST AND PASSED. ADVISED THE CUSTOMER TO CALL BACK IF ISSUES PERSIST. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301368 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 78 YR Hospitalization