UNICEL® DXC 600I SYNCHRON® ACCESS CLINICAL SYSTEM
Report
- Report Number
- 2050012-2013-00433
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- June 4, 2013
- Report Date
- June 4, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED AND DISCOVERED A FULL OVERFLOW WASTE CONTAINER AND FLUID FLOWED BACK THROUGH THE WASH STATION TUBING. THE FSE EMPTIED THE OVERFLOW WASTE CONTAINER TO RESOLVE THE LEAK. THE FSE STATED THAT A DEFECTIVE CTA (CLOSED TUBE ALIQUOTER) WASH STATION STARTED LEAKING FLUID ON (B)(4) 2013 AND THE FSE WAS DISPATCHED TO RESOLVE THE ISSUE. PLEASE REFER TO MEDWATCH REPORT # 2050012-2013-00416 FOR THE REPORT OF THE EVENT WHICH OCCURRED ON (B)(4) 2013. THE FSE REPLACED THE DEFECTIVE CTA WASH STATION TO RESOLVE THE ISSUE BUT INADVERTENTLY FAILED TO DRAIN THE WASTE OVERFLOW CONTAINER, WHICH WAS DESIGNED TO CAPTURE ANY OVERFLOW FROM THE CTA WASH STATION, DURING THE CTA WASH STATION REPLACEMENT. RESULTS: FAILURE MODE IS ATTRIBUTED TO AN OVERFILLED OVERFLOW CTA WASH STATION WASTE CONTAINER. BECKMAN COULTER INTERNAL IDENTIFIER FOR THIS REPORT IS (B)(4).
THE CUSTOMER REPORTED NOTICING A WASH BUFFER LEAK WHICH POOLED UNDERNEATH THE ALIQUOT PROBE WASH STATION INVOLVING THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM WHILE TROUBLESHOOTING LEVEL SENSE ERRORS ON THE CTA (CLOSED TUBE ALIQUOTER). THE CUSTOMER INDICATED THAT APPROXIMATELY 2 ML OF WASH BUFFER WAS PIPETTED OUT OF THE WASH TOWER AND A SMALL POOL OF FLUID HAD COLLECTED UNDERNEATH THE WASH STATION. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO ERRONEOUS PATIENT RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302622 | UNICEL® DXC 600I SYNCHRON® ACCESS CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER | NA | NA |