COULTER® AC*T DIFF ANALYZER
Report
- Report Number
- 1061932-2013-01272
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- June 6, 2013
- Report Date
- June 6, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K973634
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED AND CONFIRMED A SMALL LEAK AT THE PROBE AND THAT THE INSTRUMENT WAS GENERATING PLT (PLATELET) BACKGROUND FAILURES. THE FSE INDICATED THAT THE DUAL FILTERS REQUIRED REPLACEMENTS BUT THE CUSTOMER DID NOT HAVE ANY FILTERS IN INVENTORY. THE FSE REPLACED THE DILUENT FILTERS TO RESOLVE THE LEAK BUT DISCOVERED EXCESSIVE BUILDUP ON THE SYRINGES. THE FSE PROCEEDED TO REPLACE THE DILUENT AND SAMPLE SYRINGES AND DECONTAMINATED THE INSTRUMENT TO RESOLVE THE PLATELET BACKGROUND FAILURES. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES AND RESULTS MET PUBLISHED SPECIFICATIONS. FAILURE MODE OF THE LEAK IS ATTRIBUTED TO THE DUAL DILUENT FILTERS AND FAILURE MODE OF THE PLT BACKGROUND FAILURES IS ATTRIBUTED TO EXCESSIVE BUILDUP ON THE SYRINGES. (B)(4).
THE CUSTOMER REPORTED FAILING PLT (PLATELET) BACKGROUNDS AND NOTICED WHITE CRYSTALLIZATION NEAR THE TRIPLE SYRINGE ASSEMBLY AND DROPS OF FLUID AT THE PROBE WIPE DURING STARTUP INVOLVING THE COULTER ACT DIFF ANALYZER. THE CUSTOMER INDICATED THAT APPROXIMATELY 3 DROPS OF FLUID LEAKED AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING GLOVES AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO ERRONEOUS PATIENT RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302126 | COULTER® AC*T DIFF ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |