OLYMPUS AU481-02E CLINICAL CHEMISTRY ANALYZER (WITH ISE)
Report
- Report Number
- 9612296-2013-00085
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- June 6, 2013
- Report Date
- June 6, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- CEM
- PMA / PMN Number
- K981743
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SAMPLE COLLECTION AND CENTRIFUGATION INFORMATION WAS NOT PROVIDED. THE CUSTOMER INDICATED THAT THE K QUALITY CONTROL (QC) RECOVERED HIGH. THE OPERATOR PERFORMED AN IONIZED SELECTIVE ELECTRODE (ISE) CLEANING AND RECALIBRATED, BUT QC REMAINED HIGH. THE CUSTOMER DID NOT PROVIDE ANY DOCUMENTATION OF CALIBRATION OR QC DATA. SERVICE WAS REQUESTED. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT. THE FSE EVALUATED THE ISE AND DETERMINED THAT THE ISE REFERENCE VALVE WAS FAULTY AS THERE WERE VISIBLE AIR BUBBLES IN THE LINES DURING PRIMING. THE REFERENCE VALVE WAS REPLACED AND THE ISE SYSTEM WAS RETURNED TO OPERATIONAL. SYSTEM PERFORMANCE WAS RETURNED TO SPECIFICATIONS. SYSTEM VALIDATION WAS DOCUMENTED IN CUSTOMER QC RECORDS. THERE HAVE BEEN NO FURTHER ISSUES NOTED WITH K RECOVERY. FAILURE MODE: FAILED ISE REFERENCE VALVE.
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) IN REGARDS TO OBTAINING FALSE HIGH POTASSIUM (K) RESULTS FOR APPROXIMATELY FIFTY (50) PATIENT SAMPLES GENERATED ON THE AU481-02E CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER INDICATED THAT THE FALSE K RESULT VALUES WERE OVER 100 MMOL/L AND UPON REPEAT ON AN ALTERNATE AU INSTRUMENT GENERATED RESULTS WITHIN NORMAL RANGE. NORMAL REFERENCE RANGE FOR K IS 3.5 - 5.1 MMOL/L. NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE CUSTOMER DID NOT PROVIDE SPECIFIC PATIENT DEMOGRAPHICS OR DATA DOCUMENTATION. THERE WAS NO AFFECT TO PATIENT OR PATIENT TREATMENT IN CONNECTION TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302095 | OLYMPUS AU481-02E CLINICAL CHEMISTRY ANALYZER (WITH ISE) | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | CEM | BECKMAN COULTER | AU481-02E (WITH ISE) | NA |