UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2013-00452
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- June 4, 2013
- Report Date
- June 4, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
NO OPERATOR MAINTENANCE HAD BEEN PERFORMED PRIOR TO INCIDENT. BEC CUSTOMER TECHNICAL SUPPORT (CTS) ADVISED THE CUSTOMER TO CLEAN THE PROBE AND RE-CALIBRATE. THE CUSTOMER CLEANED THE PROBE AND INFORMED CTS THAT QUALITY CONTROL (QC) RECOVERIES WERE OUT OF SPECIFICATION AND SEVERAL TESTS WERE FAILING CALIBRATION. LEAKING WAS STILL OBSERVED. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE OBSERVED THAT THE CC SAMPLE PROBE WAS LEAKING DURING THE PRIMING CYCLE. THE FSE FOUND THE CC SAMPLE PROBE WASH COLLAR VALVE WAS NOT ALLOWING ENOUGH VACUUM. THE FSE REPLACED THE CC SAMPLE PROBE AND THE CC SAMPLE PROBE WASH COLLAR VALVE. SYSTEM PERFORMANCE WAS VERIFIED. THE FSE CONFIRMED FAILURE MODE TO BE THE CC SAMPLE PROBE WASH COLLAR VALVE. BEC IDENTIFIER FOR THIS REPORT IS (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING A CHEMISTRY CARTRIDGE (CC) SAMPLE PROBE LEAK IN THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE VOLUME OF THE LEAK WAS LESS THAN 1 ML AND WAS CONTAINED TO THE INSTRUMENT. CUSTOMER STATED BLOOD UREA NITROGEN (BUN), GLUCOSE, AND HIGH DENSITY LIPOPROTEIN (HDL) FAILED CALIBRATION FOR "SAMPLE NOT DETECTED" WHEN THE LEAK WAS DISCOVERED. THE OPERATOR STATED THAT SHE WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT DURING THIS EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302574 | UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |