FDA Adverse Event Malfunction Summary report: N

COULTER® ACT DIFF 2 ANALYZER

MDR report key: 3202943 · Received July 2, 2013

Report

Report Number
1061932-2013-01254
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
June 7, 2013
Report Date
June 7, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K990352
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONTROLS WERE RUN BEFORE THE INCIDENT, BUT NOT AFTER THE EVENT. THE UNIT CURRENTLY IS PERFORMING WITHIN QUALITY CONTROL (QC) SPECIFICATIONS WITH RESPECT TO CONTROLS (ACCURACY) AND REPRODUCIBILITY (PRECISION). THE CUSTOMER REQUESTED SERVICE AND BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE INSPECTED THE ANALYZER AND FOUND BATHS WERE NOT DRAINING. UPON FURTHER INSPECTION, THE FSE DISCOVERED THAT THE PUMP WAS NOT WORKING PROPERLY CAUSING THE BATHS NOT TO DRAIN. THE FSE REPLACED THE WASTE PUMP AND ALSO REMOVED CLOGS IN THE WASTE TUBING AND THIS RESOLVED THE ISSUE. THE FSE PERFORMED SYSTEM VERIFICATION, WHICH PASSED. FAILURE MODE WAS ATTRIBUTED TO AN INADEQUATE DRAINING OF THE BATHS DUE TO A PLUG IN THE WASTE PATHWAY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT ERRONEOUS HEMOGLOBIN (HGB) AND HEMATOCRIT (HCT) RESULTS WERE OBTAINED ON THE FIRST TWO PATIENTS AFTER PROCESSING MORNING CONTROLS ON THE COULTER ACT DIFF 2 ANALYZER. ERRONEOUS RESULTS WERE SUSPECTED BECAUSE THE HEMOGLOBIN/HEMATOCRIT (H/H) RATIO DID NOT CORRELATE. THE EXAMPLE GIVEN WAS 11 HGB WITH A 22 HCT. THE INSTRUMENT PRINTOUTS WERE REQUESTED, BUT WERE NOT PROVIDED BY THE CUSTOMER, THEREFORE FURTHER EVALUATION AND FLAGGING COULD NOT BE ASSESSED. NO ERRONEOUS RESULTS WERE REPORTED OUT THE LABORATORY. THERE WAS NO DEATH, INJURY, OR EFFECT TO PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302395 COULTER® ACT DIFF 2 ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER AC T DIFF 2 NA

Patients

Seq Age Sex Outcome Treatment
1