COULTER® LH 780 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-01248
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- June 10, 2013
- Report Date
- June 10, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061616
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE ON (B)(4) 2013. THE FSE FOUND THAT THE STARTUP WAS FAILING RETICS AND DOTTING OUT ON SAMPLES AND CONTROLS. THE RETICS OUT ON STARTUP ISSUE WAS TRACED TO A TUBING LEAK AT VALVE VL95/98. THE FSE REPLACED THE TUBING TO FIX THE LEAK. UPON FURTHER INSPECTION, THE FSE INDICATED THAT RETIC CLEAR REAGENT WOULD NOT PRIME WHICH WAS CAUSED BY A SLUGGISH ACTUATOR FOR VALVE VL95/98. THE FSE REPLACED THE ACTUATOR. THE FSE RAN THE STARTUP WITH ALL PARAMETERS NOW PASSING BACKGROUNDS. THE FSE ALSO RAN SAMPLES AND NOTED THE RETICS WERE NOT GETTING RESULTS (RETIC VOTE-OUTS, NON-NUMERIC VALUES WERE GENERATED). UPON FURTHER INVESTIGATION, THE FSE DISCOVERED THAT THE RETIC SHEAR VALVE, VL99, AND THE RETIC STAIN PUMP WERE NOT WORKING. THE FSE ORDERED PARTS AND RETURNED TO THE CUSTOMER'S LABORATORY ON (B)(4) 2013 AND REPLACED MANIFOLD (MF16). THE FSE VERIFIED THE INSTRUMENT'S OPERATIONS. FAILURE MODE WAS ATTRIBUTED TO THE TUBING BEING OFF AT THE BSV, TUBING AT VL95/98, ACTUATOR FOR VL95/98, AND 3 SOLENOIDS ON MANIFOLD (MF16) NOT FIRING. HOWEVER, THE STARTUP FAILED RETIC AND THE SAMPLES GAVE NON-NUMERIC RESULTS FOR THE RETIC PARAMETER ALERTING THE CUSTOMER TO AN INSTRUMENT PROBLEM AND TO REVIEW RESULTS. (B)(4).
THE CUSTOMER CONTACTED BECKMAN COULTER (BEC) STATING THAT WHILE RUNNING A STARTUP THIS MORNING ON THE COULTER LH 780 HEMATOLOGY ANALYZER, THE RETICS FAILED. PER CUSTOMER, WHEN THEY OPENED THE FRONT COVER THEY NOTICED A TUBING BEING OFF ON THE BLOOD SAMPLING VALVE (BSV). THE CUSTOMER INDICATED THAT THE LEAK WAS ABOUT 5 ML OF THE BLUE CLEANER SOLUTION BELOW THE BSV AREA AND WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER STATED THAT THEY RE-ATTACHED THE TUBING AND CLEANED UP THE LEAK PER THE LABORATORY'S PROTOCOL. THE CUSTOMER REPEATED THE STARTUP SEVERAL TIMES BUT WAS FAILING THE RETIC BACKGROUND, AND THEY STOPPED USING THE INSTRUMENT AT THIS TIME AND CALLED FOR SERVICE. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED BY THE CUSTOMER. THE LABORATORY¿S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THE CUSTOMER WAS WEARING GLOVES, A LABORATORY COAT AND PROTECTIVE EYE WEAR AT THE TIME OF THE EVENT. NO ONE CAME INTO DIRECT CONTACT WITH THE LEAK. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. MEDICAL ATTENTION WAS NOT SOUGHT. NO PATIENT RESULTS WERE AFFECTED AND NO ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THERE WAS NO CHANGE TO PATIENT TREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302392 | COULTER® LH 780 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH780 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |