FDA Adverse Event Summary report: N

RELIANCE SYNERGY WASHER

MDR report key: 3202878 · Received July 2, 2013

Report

Report Number
9680353-2013-00079
Date Received
July 2, 2013
Date of Event
June 4, 2013
Report Date
July 2, 2013
Manufacturer
STERIS CANADA CORPORATION
Product Code
MEC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN ARRIVED AT THE FACILITY AND FOUND A HOSE CLAMP HOLDING THE RUBBER TUBING FROM THE PUMP TO THE DRAIN HAD COME OFF, ALLOWING A LARGE AMOUNT OF WATER TO SPRAY OUT ONTO THE FLOOR. THE TECHNICIAN ATTRIBUTED THE LOOSENING OF THE HOSE CLAMP TO THE VIBRATIONS OF THE WASHER AND AGE OF THE UNIT. THE WASHER WAS INSTALLED IN MAY 2004 AND IS CURRENTLY UNDER A STERIS SERVICE CONTRACT. THE TECHNICIAN PLACED THE HOSE CLAMP BACK ONTO THE UNIT AND PROPERLY TIGHTENED THE HOSE CLAMP TO SPECIFICATION. A TEST CYCLE WAS PERFORMED AND THE UNIT WAS PLACED BACK INTO OPERATION.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THEIR RELIANCE WASHER LEAKING A LARGE AMOUNT OF WATER. THE WATER REPORTEDLY SPREAD THROUGHOUT THE ROOM AND ADJACENT HALLWAY. NO PROCEDURAL DELAYS/CANCELLATIONS OR INJURIES WERE REPORTED.