FDA Adverse Event Malfunction Summary report: N

COAGUCHEK ® XS SYSTEM

MDR report key: 3202815 · Received July 2, 2013

Report

Report Number
1823260-2013-04004
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
May 14, 2013
Report Date
July 2, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GJS
PMA / PMN Number
K062925
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. WILL NOT BE RETURNED TO MANUFACTURER.

Description of Event or Problem · 1

CALLER STATES PATIENT TESTED 4.7 INR ON THE COAGUCHEK XS SYSTEM WHILE A COMPARISON LAB RETURNED AS 3.2 INR. NO ACTIONS TAKEN BASED ON DEVICE RESULT. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT SYSTEM HOWEVER CALLER NO LONGER HAS THE TEST STRIPS; REPLACEMENT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302038 COAGUCHEK ® XS SYSTEM PROTHROMBIN TIME TEST STRIPS GJS ROCHE DIAGNOSTICS NA 21610911

Patients

Seq Age Sex Outcome Treatment
1 COUMADIN