ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-09551
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Report Date
- June 6, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: ON INVESTIGATION, THE PUMP POWERED ON WITH A DIM, DISCOLORED DISPLAY. A TEST DISPLAY WAS ATTACHED TO THE PUMP AND ILLUMINATED PROPERLY AND WAS CLEARLY LEGIBLE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. REPORTEDLY, THE DISPLAY WAS GRADUALLY FADING AND CAN ONLY BE SEEN IN A DARK ROOM. IT WAS NOTED THAT THE LETTERING WAS DISCOLORED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.