FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3202492 · Received July 2, 2013

Report

Report Number
2955842-2013-02394
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
May 28, 2013
Report Date
June 10, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED COMPLAINT. THE INSTRUMENT WAS PLACED ON AN IN-HOUSE IS3000 SYSTEM AND WAS DRIVEN. THE INSTRUMENT PASSED RECOGNITION AND ENGAGEMENT TESTING. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIPS OPENED AND CLOSED PROPERLY. THE FUNCTIONAL PERFORMANCE TESTING DID NOT FIND ANY ISSUES. ENGINEERING ALSO FOUND TUBE DAMAGE. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .090 - .175 IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY CAUSED BY MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE THE MARYLAND BIPOLAR FORCEPS INSTRUMENT TIPS WERE ALLEGEDLY NOT ARTICULATING. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301328 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 M11130319 966

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES