FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60

MDR report key: 3202478 · Received July 2, 2013

Report

Report Number
3005075853-2013-03384
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
June 12, 2013
Report Date
June 13, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

THE ANALYSIS FOUND THAT ONE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE; THE DEVICE CLOSED, FIRED AND OPENED WITHOUT ANY DIFFICULTIES NOTED. IT SHOULD BE NOTED THAT IN ORDER TO OPEN A DEVICE THAT HAS BEEN PARTIALLY FIRED OR FULLY FIRED A REVERSE STROKE NEEDS TO BE PERFORMED TRIGGER TO TRIGGER TO HANDLE IN ORDER TO RETURN THE KNIFE TO THE HOME POSITION (INDICATOR IN THE "0" POSITION) AND PRESS THE ANVIL RELEASE BUTTON TO OPEN. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW ANTERIOR COLECTOMY PROCEDURE, THE DEVICE LOCKED AND WOULD NOT OPEN AFTER THE SEVENTH FIRING. THE DEVICE WAS NOT ON TISSUE. ALL FIRING WERE WITH WHITE CARTRIDGES. NO ISSUE NOTED WITH PREVIOUS FIRINGS. IT IS UNKNOWN ON WHAT TISSUE THE DEVICE WAS FIRED. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302182 ECHELON*FLEX60 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA K4CL2L

Patients

Seq Age Sex Outcome Treatment
1 ECR60W