PINNACLE 100 ACET CUP 56MM

MDR report key: 3202444 · Received July 2, 2013

Report

Report Number
1818910-2013-20396
Event Type
Injury
Date Received
July 2, 2013
Date of Event
February 12, 2013
Report Date
June 17, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
JDI
PMA / PMN Number
PK001534
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND NO ADDITIONAL RELATED REPORTS FOR THE LOT CODES 2293504, 2176469, 2122168, AND BA3D11000. A SEARCH OF THE COMPLAINT DATABASE FINDS ADDITIONAL REPORTED INCIDENTS AGAINST LOT CODES 2291367 AND A53FX1000 SINCE ITS RELEASE FOR DISTRIBUTION; HOWEVER, A REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THE A53FX1000 AND 2291367 LOT CODES DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE REMAINING PRODUCT AND LOT COMBINATIONS. MEDICAL RECORDS AND X-RAY(S) WERE REVIEWED. WITH THE INFORMATION PROVIDED, THE EVIDENCE SUGGESTS THE PATIENT HAD AN ADVERSE REACTION TO METAL DEBRIS AND METALLOSIS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THE A53FX1000 AND 2291367 LOT CODES DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE REMAINING PRODUCT AND LOT COMBINATIONS. MEDICAL RECORDS AND X-RAY(S) WERE REVIEWED BY A MEDICAL PROFESSIONAL. WITH THE INFORMATION PROVIDED, THE EVIDENCE SUGGESTS THE PATIENT HAD AN ADVERSE REACTION TO METAL DEBRIS AND METALLOSIS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

UDI: (B)(4).

Description of Event or Problem · 1

PATIENT WAS REVISED BILATERALLY TO ADDRESS OSTEOLYSIS AND MASSIVE EFFUSION GREATER THAN 50CC. **UPDATE** (B)(4) 2013 - CLINICAL REPORT STATES THE PATIENT WAS REVISED TO ADDRESS METALLOSIS AND ELEVATED COBALT CHROME LEVELS (BILATERALLY). THE ASSOCIATED EMDR'S WERE UPDATED. NO CHANGE IN THE EXISTING MDR DECISION. **UPDATE: (B)(4) 2013 - LITIGATION PAPERS RECEIVED. LITIGATION ALLEGES PAIN, METALLOSIS, FLUID BUILD UP AND EROSION OF THE ACETABULUM. THE ACETABULAR CUP IS NOW BEING REPORTED TO ADDRESS THE EROSION.

Description of Event or Problem · 1

PPF ALLEGES METAL WEAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302083 PINNACLE 100 ACET CUP 56MM ACETABULAR CUP JDI DEPUY ORTHOPAEDICS INC US A53FX1000

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention