FDA Adverse Event Malfunction Summary report: N

SYNCHROMED II

MDR report key: 3202377 · Received July 2, 2013

Report

Report Number
3004209178-2013-00090
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
March 5, 2012
Report Date
March 29, 2012
Manufacturer
MDT PUERTO RICO OPERATIONS CO
Product Code
LKK
PMA / PMN Number
P860004
Removal / Correction Number
Z-0591-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: PRODUCT ID: 8709, SERIAL# (B)(4), IMPLANTED: (B)(6) 2005, PRODUCT TYPE: CATHETER. PRODUCT ID: 8598A, SERIAL# (B)(4), IMPLANTED: (B)(6) 2011, PRODUCT TYPE: CATHETER. PRODUCT ID: 8596, SERIAL# (B)(4), IMPLANTED: (B)(6) 2005, PRODUCT TYPE: CATHETER. PRODUCT ID: 8590-8, LOT# N264695, IMPLANTED: (B)(6) 2011,PRODUCT TYPE: ACCESSORY. (B)(4).

Description of Event or Problem · 1

A MOTOR STALL WAS REPORTED. THE PATIENT HAD A MAGNETIC RESONANCE IMAGING (MRI) SCAN AT AN UNKNOWN DATE AND DID NOT HAVE THE PUMP INTERROGATED AFTERWARDS. THE PATIENT EXPERIENCED ¿SOME PAIN¿ AND HAD BEEN TAKING ORAL MEDICATION. ON THE DATE OF THIS REPORT, THE PATIENT CAME IN FOR A REFILL AT WHICH TIME THE PUMP WAS INTERROGATED. THE PUMP¿S EVENT LOGS NOTED THAT A MOTOR STALL OCCURRED ON (B)(6) 2012 IT SAID, ¿STOPPED PUMP PERIOD MAY EXCEED TUBE SET.¿ A CRITICAL ALARM WAS HEARD. A ¿MOTOR STALL RECOVERY OCCURRED¿ MESSAGE WAS SEEN ON THE DATE AND AT THE TIME OF PUMP INTERROGATION. THE DRUGS IN THE PUMP WERE DILAUDID (HYDROMORPHONE) AND BUPIVACAINE. ADDITIONAL INFORMATION CONFIRMED THAT THE PUMP STALLED DURING THE MRI. IT WAS NOTED THAT THE PATIENT DID NOT EXHIBIT ANY CLINICAL SYMPTOMS DURING THIS TIME. THE PATIENT HAD BEEN BACK FOR REFILLS SINCE THIS INCIDENT AND EVERYTHING HAD BEEN APPROPRIATE IN TERMS OF VOLUMES IN THE PUMP. THE PATIENT¿S PAIN RELIEF WAS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301930 SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MDT PUERTO RICO OPERATIONS CO 863720

Patients

Seq Age Sex Outcome Treatment
1 00056 YR