FDA Adverse Event Death Summary report: N

INFINIA HAWKEYE 4

MDR report key: 3202355 · Received June 13, 2013

Report

Report Number
3202355
Event Type
Death
Date Received
June 13, 2013
Date of Event
June 5, 2013
Report Date
June 11, 2013
Manufacturer
GENERAL ELECTRIC (GE) HEALTHCARE
Product Code
KPS
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

ON (B)(6) 2013, WHILE A (B)(6) PT WAS UNDERGOING AN INDIUM111 OCTREOSCAN (FULL BODY SCAN), THE TOP DETECTOR OF THE GA INFINIA HAWKEYE 4 SPECT/CT GAMMA CAMERA EXPERIENCED A CATASTROPHIC MECHANICAL FAILURE. PRIOR TO THE PT SCAN, THE GE INFINIA HAWKEYE 4 SPECT/CT GAMMA CAMERA PASSED A SELF TEST CHECK WITHOUT ANY PROBLEMS. THE PT WAS POSITIONED ON THE TABLE. THE SCAN STARTED. THE TABLE STARTED TO MOVE SLOWLY, AS EXPECTED, INTO THE STARTING POSITION. THE TABLE WAS SET WITH THE PT'S HEAD UNDER THE TOP DETECTOR. AS EXPECTED, THE TOP DETECTOR MOVED DOWN AND STOPPED AN INCH FROM THE PT. THE TECH STARTED THE SCAN ACQUIRING IMAGES. APPROX ONE MIN LATER, THE TOP DETECTOR FELL ON TOP OF THE PT'S HEAD AND UPPER ABDOMEN. IT THEN LOCKED INTO PLACE. THE DETECTOR WEIGHED APPROX 650 TO 800 POUNDS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
269849 INFINIA HAWKEYE 4 SPECT/CT GAMMA CAMERA KPS GENERAL ELECTRIC (GE) HEALTHCARE INFINIA HAWKEYE 4

Patients

Seq Age Sex Outcome Treatment
1 66 YR Death