ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2013-00280
- Event Type
- Death
- Date Received
- June 24, 2013
- Date of Event
- May 23, 2013
- Report Date
- May 28, 2013
- Manufacturer
- CAREFUSION CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K950419
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). LOGS RECEIVED ONLY, DEVICE RECEIPT PENDING. AN EVENT LOG REVIEW WAS PERFORMED; NO PROGRAMMING ISSUES WERE IDENTIFIED THAT WOULD EXPLAIN WHY THE DOSE COMPLETED EARLY. DEVICE WERE REQUESTED. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVAL HAS BEEN COMPLETED.
THE CUSTOMER REPORTS ZOSYN 100ML WAS TO BE INFUSED OVER 4 HOURS, BUT THE COMPLETE DOSE WAS INFUSED IN 20 MINUTES. THE NURSE STATE THAT SHE PROGRAMMED IT FOR A "REGULAR INFUSION" AND NOT AND "EXTENDED INFUSION." FOUR PC UNITS AND 14 PUMP MODULES WERE IN USE AND THE CUSTOMER IS NOT SURE WHICH PUMP WAS INFUSING THE ZOSYN AND CANNOT MATCH THE DATE AND TIME THE EVENT WAS SUPPOSED TO HAVE HAPPENED. WITH THE LOGS FROM THE PUMP. THE CUSTOMER STATED THAT PT WENT TO SURGERY AND DIED LATER; HOWEVER, "THE CAUSE OF DEATH WAS MASSIVE RIGHT HEART FAILURE AND IN NO MEANS RELATE TO THE ZOSYN INFUSION." A SIMILAR EVENT OCCURRED ON THE SAME PUMP MODULE LATER IN THE DAY. SEE REPORT #2016493-2013-00286 FOR THE SECOND EVENT. NO FURTHER PT OR EVENT INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 284769 | ALARIS PUMP MODULE | FRN | CAREFUSION CORPORATION | 8100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Death | ALARIS PUMP MODULE: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PC UNIT: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PC UNIT: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ECMO, MFR UNK: SERIAL # UNK| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PC UNIT: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4)| ALARIS PC UNIT: SN (B)(4)| ALARIS PUMP MODULE: SN (B)(4) |