FDA Adverse Event
Malfunction
Summary report: N
TERUMO ADVANCED PERFUSION SYSTEM 1
MDR report key: 3202214
·
Received June 28, 2013
Report
- Report Number
- 1828100-2013-00610
- Event Type
- Malfunction
- Date Received
- June 28, 2013
- Date of Event
- June 7, 2013
- Report Date
- June 7, 2013
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DTQ
- PMA / PMN Number
- K022947
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
EVALUATION IS IN PROGRESS, BUT NOT YET CONCLUDED. THE FIELD SERVICE REPRESENTATIVE (FSR) RETURNED THE SUSPECT BATTERIES, POWER MANAGER BOARD AND DATA LOGS TO THE MFR FRO FURTHER EVALUATION. THE DATA LOG ANALYSIS CONFIRMED THE REPORTED ISSUE.
Description of Event or Problem · 1
THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE PERFUSION SYSTEM WOULD NOT BOOT UP ON BATTERY, AS THE BATTERIES WERE DEAD. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 295932 | TERUMO ADVANCED PERFUSION SYSTEM 1 | DTQ | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 801763 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |