FDA Adverse Event Malfunction Summary report: N

NEWTON IQ W/STAY SAFE CYCLER SET

MDR report key: 3202172 · Received June 28, 2013

Report

Report Number
8030665-2013-00397
Event Type
Malfunction
Date Received
June 28, 2013
Date of Event
June 3, 2013
Report Date
June 3, 2013
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
KDJ
PMA / PMN Number
K904706
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONLY THE BLUE PUMP TUBING CONNECTED TO A SECTION OF DRAIN LINE WAS RETURNED TO THE MFR FOR PHYSICAL EVALUATION. A FUNCTIONAL TEST WAS PERFORMED AND NO LEAKS WERE DETECTED. AN INVESTIGATION OF THE DEVICE MFG RECORDS WAS CONDUCTED BY THE MFR. THERE WERE NO DEVIATIONS OR NONCONFORMANCE DURING THE MFG PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.

Description of Event or Problem · 1

THE PT ALLEGED A LEAK OCCURRED AT THE CONNECTION OF THE DRAIN LINE AND BLUE PUMP TUBING. THERE WAS NO FLUID IN, ON OR NEAR THE CYCLER. LEAK WAS FOUND AFTER TREATMENT WAS COMPLETED. THE PT'S PD RN STATED WHO STATED THE PT NEEDED NO MEDICAL INTERVENTION. EFFLUENT REMAINS CLEAR. SAMPLE IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
296614 NEWTON IQ W/STAY SAFE CYCLER SET KDJ FRESENIUS MEDICAL CARE NORTH AMERICA 13AR08875

Patients

Seq Age Sex Outcome Treatment
1 63 YR NEWTON CYCLER