FDA Adverse Event
Malfunction
Summary report: N
NEWTON IQ W/STAY SAFE CYCLER SET
MDR report key: 3202172
·
Received June 28, 2013
Report
- Report Number
- 8030665-2013-00397
- Event Type
- Malfunction
- Date Received
- June 28, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 3, 2013
- Manufacturer
- FRESENIUS MEDICAL CARE NORTH AMERICA
- Product Code
- KDJ
- PMA / PMN Number
- K904706
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ONLY THE BLUE PUMP TUBING CONNECTED TO A SECTION OF DRAIN LINE WAS RETURNED TO THE MFR FOR PHYSICAL EVALUATION. A FUNCTIONAL TEST WAS PERFORMED AND NO LEAKS WERE DETECTED. AN INVESTIGATION OF THE DEVICE MFG RECORDS WAS CONDUCTED BY THE MFR. THERE WERE NO DEVIATIONS OR NONCONFORMANCE DURING THE MFG PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.
Description of Event or Problem · 1
THE PT ALLEGED A LEAK OCCURRED AT THE CONNECTION OF THE DRAIN LINE AND BLUE PUMP TUBING. THERE WAS NO FLUID IN, ON OR NEAR THE CYCLER. LEAK WAS FOUND AFTER TREATMENT WAS COMPLETED. THE PT'S PD RN STATED WHO STATED THE PT NEEDED NO MEDICAL INTERVENTION. EFFLUENT REMAINS CLEAR. SAMPLE IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 296614 | NEWTON IQ W/STAY SAFE CYCLER SET | KDJ | FRESENIUS MEDICAL CARE NORTH AMERICA | 13AR08875 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | NEWTON CYCLER |