FDA Adverse Event
Malfunction
Summary report: N
LIBERTY CYCLER SET, SINGEL CONN./EXT.DL
MDR report key: 3202141
·
Received June 28, 2013
Report
- Report Number
- 8030665-2013-00404
- Event Type
- Malfunction
- Date Received
- June 28, 2013
- Date of Event
- June 1, 2013
- Report Date
- June 1, 2013
- Manufacturer
- REYNOSA MANUFACTURING
- Product Code
- FKX
- PMA / PMN Number
- K043363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION AND THE FAILURE MODE CANNOT BE CONFIRMED. HOWEVER, AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NONCONFORMANCE'S DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.
Description of Event or Problem · 1
A PERITONEAL DIALYSIS PT HAS REPORTED THAT DIALYSIS SOLUTION WAS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLER. PT DISCOVERED FLUID INSIDE THE CASSETTE DOOR AND HEAD PUMP AREA WHEN REMOVING THE TUBING SET. LEAK LOCATION COULD NOT BE IDENTIFIED. PT HAD NO ADVERSE EFFECTS. SAMPLE HAS NOT BEEN RETURNED TO THE MANUFACTURER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 296606 | LIBERTY CYCLER SET, SINGEL CONN./EXT.DL | FKX | REYNOSA MANUFACTURING | 13CR08903 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LIBERTY DIALYSIS CYCLER |