FDA Adverse Event Malfunction Summary report: N

HEARTMATE II LVAS

MDR report key: 3202043 · Received June 27, 2013

Report

Report Number
2916596-2013-00796
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
June 3, 2013
Report Date
June 3, 2013
Manufacturer
THORATEC CORP.
Product Code
DSQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT REMAINS ON LVAD SUPPORT WITH THE PUMP. THE DEVICE REMAINS IMPLANTED AND IN USE BY THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). IT WAS REPORTED DURING IMPLANTATION OF THE PUMP, THAT THE PUMP AUTOMATICALLY TURNED OFF WHILE DE-AIRING. THE PUMP SPEED WAS CHANGED TO 7000 RPM AND THE PUMP CONTINUED TO STOP AUTOMATICALLY. THE SYSTEM CONTROLLER WAS EXCHANGED HOWEVER THERE WERE NO IMPROVEMENTS. ECHOCARDIOGRAM SHOWED NO AIR IN THE LEFT VENTRICLE. WHEN THE PUMP SPEED WAS INCREASED TO 8000 RPM, THE PUMP WAS ABLE TO START AND A LOT OF AIR WAS PRESENT IN THE LEFT VENTRICLE. THE PUMP OUTFLOW HAD TO BE DISCONNECTED IN ORDER TO GET RID OF ALL THE FROTHY BLOOD DUE TO AIR IN THE PUMP. SOME AIR GOT DOWN TO THE RIGHT CORONARY ARTERY AND CAUSED TRANSIENT ST ELEVATION. CENTRAL VENOUS PRESSURE (CVP) WAS OKAY INDICATING THE RIGHT VENTRICLE FUNCTION IS STILL ACCEPTABLE. THE HOSPITAL TEAM CONTINUED TO DE-AIR THE PUMP UNTIL NO MORE AIR COULD BE SEEN UNDER AN ECHOCARDIOGRAM. THE FINAL SPEED WAS SET AT 8800 RPM WITH LOW SPEED LIMIT AT 8400 RPM; FLOW WAS AROUND 3-4 WITH FREQUENT "---" DISPLAYED. THE PATIENT WOKE UP AND WAS EXTUBATED 10 HOURS POST OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
292177 HEARTMATE II LVAS DSQ: LEFT VENTRICULAR ASSIST DEVICE DSQ THORATEC CORP. 104911 125085

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention