FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3201940 · Received June 27, 2013

Report

Report Number
8020893-2013-01398
Event Type
Malfunction
Date Received
June 27, 2013
Date of Event
May 30, 2013
Report Date
May 30, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK SCREEN. THERE WAS NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND REPLACED THE PRINTED WIRING BOARD (PWB) PCB AND DISPLAY CABLE. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291893 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1