FDA Adverse Event Malfunction Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3201786 · Received July 2, 2013

Report

Report Number
2024168-2013-04139
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
June 7, 2013
Report Date
June 7, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4) - INCORRECT PREPARATION. IT WAS INITIALLY REPORTED THAT THE DEVICE WOULD BE RETURNED FOR ANALYSIS. SUBSEQUENT INFORMATION REVEALED THAT THE DEVICE WAS DISCARDED AND IS NOT AVAILABLE FOR EVALUATION. IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. IT SHOULD BE NOTED THAT THE XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM INSTRUCTIONS FOR USE INSTRUCTS THE PHYSICIAN TO REMOVE THE PROTECTIVE SHEATH AND STYLET PRIOR TO PREPPING THE DEVICE FOR USE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS) WAS REMOVED FROM THE PLASTIC HOOP WITHOUT RESISTANCE. THE SDS WAS PREPARED WHILE THE SHEATH WAS ON THE BALLOON, AS THIS IS THE NORMAL PRACTICE OF THE LAB. THE SHEATH AND STYLET WERE REMOVED WITHOUT ANY RESISTANCE AND THE STENT CAME OFF THE BALLOON. THERE WAS NO PATIENT INVOLVEMENT. ANOTHER XIENCE XPEDITION WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303018 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2082261

Patients

Seq Age Sex Outcome Treatment
1