FDA Adverse Event Injury Summary report: N

840 VENTILATOR

MDR report key: 3201744 · Received June 28, 2013

Report

Report Number
8020893-2013-01431
Event Type
Injury
Date Received
June 28, 2013
Date of Event
May 29, 2013
Report Date
May 29, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

RECEIVED INFO STATING THAT DUE TO A VENTILATOR MALFUNCTION PT WAS PLACED ON A SECOND VENTILATOR. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE INSPIRATORY PCB AND KEYBOARD. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294915 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention