FDA Adverse Event Injury Summary report: N

840 VENTILATOR

MDR report key: 3201734 · Received June 28, 2013

Report

Report Number
8020893-2013-01428
Event Type
Injury
Date Received
June 28, 2013
Date of Event
June 3, 2013
Report Date
June 28, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

COVIDIEN RECEIVED INFO STATING THAT DUE TO A MALFUNCTION OF THE 840 VENTILATOR. THE PT WAS PLACED ON A SECOND VENTILATOR. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE OXYGEN SENSOR. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
295813 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention