FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3201592 · Received June 26, 2013

Report

Report Number
2518422-2013-01307
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
May 31, 2013
Report Date
May 31, 2013
Manufacturer
RESPIRONICS INC., SLEEP & HOME RESPIRATORY GROUP
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MFR FOR ROUTINE PREVENTATIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291368 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC., SLEEP & HOME RESPIRATORY GROUP 1054260

Patients

Seq Age Sex Outcome Treatment
1