FDA Adverse Event Malfunction Summary report: N

TRILOGY O2

MDR report key: 3201551 · Received June 26, 2013

Report

Report Number
2518422-2013-01291
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
June 5, 2013
Report Date
June 5, 2013
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S OXYGEN BLENDING MODULE BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MFR RECEIVED INFO ALLEGING A "SERVICE REQUIRED" ALARM CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291389 TRILOGY O2 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1040004