FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3201387 · Received June 26, 2013

Report

Report Number
1720753-2013-07622
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
June 20, 2013
Report Date
June 26, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND ADJUSTED THE HIGH VOLTAGE SUPPLY REGULATOR. THE SYSTEM OPERATES AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED AN "MA SENSOR FAILED" ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290970 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1