FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3201247 · Received July 2, 2013

Report

Report Number
2531779-2013-09495
Event Type
Malfunction
Date Received
July 2, 2013
Report Date
June 4, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 08/21/2013 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/29/2013 WITH THE FOLLOWING FINDINGS:THE KEYPAD WAS FOUND TO BE FULLY INTACT; THERE WAS NO PEELING OR DAMAGE OBSERVED. DURING TESTING, THE OK KEYPAD BUTTON WAS FOUND TO BE INTERMITTENTLY UNRESPONSIVE; ALL OTHER KEYPAD BUTTONS RESPOND APPROPRIATELY. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER CLAIMED THE UP KEYPAD BUTTON WAS HARD TO PUSH AND INTERMITTENTLY UNRESPONSIVE WHEN PRESSED. AT THE TIME OF TROUBLESHOOTING, THE REPORTER DENIED ANY VISIBLE DAMAGE TO THE KEYPAD. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. .

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303135 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 9 YR