FDA Adverse Event Injury Summary report: N

DRILL 3334800 VISAO 80K HANDPIECE

MDR report key: 3201239 · Received July 2, 2013

Report

Report Number
1045254-2013-00440
Event Type
Injury
Date Received
July 2, 2013
Date of Event
June 10, 2013
Report Date
June 10, 2013
Manufacturer
MEDTRONIC XOMED, INC.
Product Code
ERL
PMA / PMN Number
K011321
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(6). UPDATED EVENT DESCRIPTION: THIS PATIENT IS FEMALE WHO COMPLAINS OF A HISTORY OF A PALATINE TORUS. DURING THE CASE, PER THE SURGEON, THE SHAFT OF THE DRILL ABOVE THE DRILL BUR GENERATED SIGNIFICANT HEATING CAUSING A RIGHT BUCCAL MUCOSAL BURN THAT WAS SECOND DEGREE. BACITRACIN OINTMENT WAS APPLIED TO THIS AREA. THE PATIENT WAS DISCHARGED HOME. IT HAS BEEN CONFIRMED THAT THE PATIENT IS STILL HEALING. (B)(4).

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. CONCOMITANT PRODUCT: 3334635, GUARD 3334635 VISAO BUR W/IRRIGATION, LOT AND S/N NOT PROVIDED. ERL 510K: K983224. (B)(4). NO EVALUATION SUMMARY AVAILABLE, DEVICE NOT RETURNED FOR EVALUATION.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

THIS PATIENT IS FEMALE WHO COMPLAINS OF A HISTORY OF A PALATINE TORUS. DURING THE CASE, PER THE SURGEON, THE SHAFT OF THE DRILL ABOVE THE DRILL BUR GENERATED SIGNIFICANT HEATING CAUSING A RIGHT BUCCAL MUCOSAL BURN THAT WAS SECOND DEGREE. BACITRACIN OINTMENT WAS APPLIED TO THIS AREA. THE PATIENT WAS DISCHARGED HOME. IT HAS BEEN CONFIRMED THAT THE PATIENT IS STILL HEALING.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT SUSTAINED AN ORAL MUCOSAL BURN WHEN THE DRILL WAS BEING USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301542 DRILL 3334800 VISAO 80K HANDPIECE DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL MEDTRONIC XOMED, INC. 3334800 71603100

Patients

Seq Age Sex Outcome Treatment
1 00059 YR Other