FDA Adverse Event Malfunction Summary report: N

UV FLASH TRANSFER SETS

MDR report key: 3201191 · Received July 2, 2013

Report

Report Number
1416980-2013-16998
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
June 11, 2013
Report Date
June 11, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN EVENT INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE UV FLASH TRANSFER SETS WAS IDENTIFIED. A SAMPLE OF THE DEVICE WAS RECEIVED AND IS PENDING EVALUATION. A BATCH REVIEW WAS CONDUCTED FOR POTENTIALLY ASSOCIATED LOT NUMBERS H12K26058 AND NO EXCEPTIONS WERE OBSERVED THAT WERE RELATED TO THE REPORTED CONDITION. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). AN EVALUATION OF THE ACTUAL SAMPLE WAS CONDUCTED. VISUAL INSPECTION, LEAK TESTING AND CLEAR PASSAGE TESTING WAS PERFORMED WITH NO ISSUES NOTED. FUNCTIONALLY HAND TIGHTENED A LAB TITANIUM ADAPTER TO SET WITH DIFFICULTY NOTED. SAMPLE WAS CONFIRMED FOR THE REPORTED PROBLEM. DIMENSIONAL INSPECTION OF THE RETURNED TRANSFER SET SAMPLE IDENTIFIED THAT THE INSIDE DIAMETER OF THE CATHETER ADAPTER SHROUD WAS BELOW NOMINAL. IMPROVEMENTS WERE MADE TO THE MOLD AND MOLDING DEPARTMENT PROCEDURES. A FOLLOW UP REPORT WILL BE FILED IF ANY ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THIS IS REPORT 2 OF 3 INVOLVED IN THIS EVENT. THIS IS A REPORT OF A HOME PATIENT (HP) WHO HAD DIFFICULTY CONNECTING THE TRANSFER SET TO THE TITANIUM ADAPTER FOR USE DURING DIANEAL THERAPY. IT WAS REPORTED THAT A 3 MM GAP WAS OBSERVED AT THE JUNCTION OF THE DEVICES. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301801 UV FLASH TRANSFER SETS SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H12K26058

Patients

Seq Age Sex Outcome Treatment
1 LOCKING TITANIUM ADAPTER