FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT ?

MDR report key: 3201178 · Received July 2, 2013

Report

Report Number
2134265-2013-04533
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
March 28, 2013
Report Date
June 12, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE IS A COMBINATION PRODUCT. (B)(4). DEVICE EVALUATED BY MANUFACTURER - A VISUAL AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE NOTED THAT THE STENT WAS DAMAGED. THE ENTIRE LENGTH OF THE STENT WAS SEVERELY STRETCHED AND DAMAGED AND AS A RESULT OF THIS THE STENT EXTENDED APPROXIMATELY 15 MM OVER THE TIP OF THE DEVICE. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. NO OTHER ISSUES WERE NOTED WITH THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED (B)(4) 2013. IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY, THE DEVICE WAS UNABLE TO CROSS THE LESION. THE TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). THE 2.25X20MM PROMUS ELEMENT STENT WAS ADVANCED HOWEVER WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER 2.25X20MM PROMUS ELEMENT STENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS STABLE, HOWEVER RETURNED DEVICE ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301633 PROMUS ELEMENT ? STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911320220 15328549

Patients

Seq Age Sex Outcome Treatment
1