FDA Adverse Event Summary report: N

BRILLIANCE CT 16-SLICE (AIR)

MDR report key: 3201142 · Received June 28, 2013

Report

Report Number
1525965-2013-00186
Date Received
June 28, 2013
Report Date
June 3, 2013
Manufacturer
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
Product Code
JAK
PMA / PMN Number
K012009
Removal / Correction Number
1525965-06/29/2012-013-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NOTE: WE HAVE NOT COMPLETED OUR INVESTIGATION OF THIS EVENT. WE WILL FILE A FOLLOW-UP MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4).

Description of Event or Problem · 1

PHILIPS RECEIVED INFORMATION FROM A CUSTOMER SITE THAT DURING THE IMPLEMENTATION OF (B)(4) (J-BRACKETS BOLTS REPLACEMENT), THE PHILIPS FIELD SERVICE ENGINEER (FSE) FOUND 1 OF THE 16 J-BRACKET BOLTS HAD THE HEAD BROKEN OFF. THERE WAS NO REPORT OF HARM TO A PATIENT OR OPERATOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
295467 BRILLIANCE CT 16-SLICE (AIR) JAK PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. 728246

Patients

Seq Age Sex Outcome Treatment
1