ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-09474
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Report Date
- June 25, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 10/23/2013 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/11/2013 WITH THE FOLLOWING FINDINGS:THE PUMP POWERED ON TO THE VERIFY SCREEN WITH APPROPRIATELY AUDIBLE TONES AND VIBRATION. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ALARMS OR OTHER ISSUES DUPLICATED. A 10 UNIT NORMAL BOLUS AND A 10 UNIT AUDIO BOLUS WERE PERFORMED AND WERE RECORDED ACCURATELY IN THE PUMP HISTORY. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS FOUND TO BE DISCOLORED. THE DISPLAY SCREEN WAS REPLACED WITH A TEST SCREEN AND THE PUMP DISPLAY RETURNED TO NORMAL. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PATIENT CONTACTED ANIMAS ON (B)(6) 2013 RELATED TO A LOSS OF PRIME ON THE PUMP. DURING TROUBLESHOOTING OF THE LOSS OF PRIME, ONE BOLUS ENTRY WAS NOTED TO BE MISSING IN THE PUMP BOLUS HISTORY. THE PATIENT INDICATED THAT A BOLUS WAS DELIVERED AT 6:31 PM ON (B)(6) 2013 AND ANOTHER WAS AT 6:51 PM BUT THERE WAS ONE BOLUS IN BETWEEN THAT WAS NOT LISTED IN THE BOLUS HISTORY. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE PUMP BOLUS HISTORY ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 303150 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |