FDA Adverse Event Malfunction Summary report: N

AVS PL SPACER 8 X 25 X 4 DEG - 8

MDR report key: 3200964 · Received July 2, 2013

Report

Report Number
0009617544-2013-00251
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
April 30, 2013
Report Date
June 5, 2013
Manufacturer
STRYKER SPINE-FRANCE
Product Code
MQP
PMA / PMN Number
K050624
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF MADE AVAILABLE WILL BE REPORTED IN A SUPPLEMENTAL REPORT. METHOD, RESULT, AND CONCLUSION CODES WILL BE MADE AVAILABLE FOLLOWING AN ENGINEERING EVALUATION.

Additional Manufacturer Narrative · 1

METHOD: DEVICE HISTORY REVIEW, COMPLAINT HISTORY REVIEW, RISK ASSESSMENT. RESULTS: THE CUSTOMER REPORTED A FRACTURE OF AN AVS PL SPACER. THE DEVICE WAS NOT RETURNED FOR EVALUATION. MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE IDENTIFIED. IT IS LIKELY THAT THE USER MAY HAVE EXCESSIVELY IMPACTED THE SPACER WITH THE MALLET DURING INSERTION. NO FURTHER EVALUATION POSSIBLE. CONCLUSION: THE ROOT CAUSE OF THE FAILURE COULD NOT BE DETERMINED DUE TO THE ABSENCE OF THE DEVICE AND LIMITED INFORMATION.

Description of Event or Problem · 1

"AS THE SURGEON WAS PLACING PEEK CAGE IN AT LEFT L5-S1 WITH INSERTER AND MALLET, HE NOTED THE PEEK CAGE HAD BROKEN INTO TWO (2) PIECES. BOTH PIECES WERE REMOVED AND NEW PEEK CAGE INSERTED. NO PATIENT HARM." INTENDED PROCEDURE WAS "TRANSFORAMINAL LUMBAR INTERBODY FUSION AT L5-S1 USING PEEK CAGES AND VITOSS PUTTY."

Description of Event or Problem · 1

"AS THE SURGEON WAS PLACING PEEK CAGE IN AT LEFT L5-S1 WITH INSERTER AND MALLET, HE NOTED THE PEEK CAGE HAD BROKEN INTO TWO (2) PIECES. BOTH PIECES WERE REMOVED AND NEW PEEK CAGE INSERTED. NO PATIENT HARM." INTENDED PROCEDURE WAS "TRANSFORAMINAL LUMBAR INTERBODY FUSION AT L5-S1 USING PEEK CAGES AND VITOSS PUTTY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303529 AVS PL SPACER 8 X 25 X 4 DEG - 8 IMPLANT MQP STRYKER SPINE-FRANCE 66320

Patients

Seq Age Sex Outcome Treatment
1