FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3200942 · Received July 2, 2013

Report

Report Number
2531779-2013-09469
Event Type
Malfunction
Date Received
July 2, 2013
Report Date
June 6, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/06/2013 WITH THE FOLLOWING FINDINGS: THERE WAS NO VISIBLE DAMAGE TO THE KEYPAD; THE KEYPAD BUTTON SYMBOLS WERE FOUND TO BE WORN. THE OK KEYPAD BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONSIVE AND CAUSED SCROLLING WHEN PRESSED. THE KEYPAD COVER WAS REMOVED; CONTAMINATION WAS FOUND UNDER THE BUTTON KEY CONTACTS. UNRELATED TO THE COMPLAINT, THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED EXTENDING FROM THE THREADS TO THE FINGER PAD, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE USER GUIDE WARNS THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND/OR THE WATERPROOF FEATURE OF THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE; UP AND DOWN ARROW BUTTONS INTERMITTENTLY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303266 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 51 YR