FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3200904 · Received July 2, 2013

Report

Report Number
2531779-2013-09451
Event Type
Malfunction
Date Received
July 2, 2013
Report Date
June 6, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED MULTIPLE CS 064-0001 ALARMS WITH HIGH FORCE OBSERVED IN THE BLACK BOX DUE TO OCCLUSION DURING PRIME. TIME AND DATE WAS ACCURATELY SET AT START OF INVESTIGATION. INVESTIGATORS PERFORMED PUMP REWIND, LOAD, AND PRIME WITH NO ALARMS. THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE WITH NO ALARMS DUPLICATED. THE PUMP WAS OPENED FOR FURTHER INVESTIGATION, INSPECTED THE MOTOR ASSEMBLY AND TESTED THE INTERNAL MOTOR NO DEFECT FOUND OR NO MOISTURE CORROSION WAS OBSERVED. UNRELATED TO THE COMPLAINT, THE PUMP BOOTED TO THE VERIFY SCREEN WITH DIM PINK CONTRAST. THE DISPLAY LENS WAS ALSO FOUND TO BE SCRATCHED. UNRELATED TO THE COMPLAINT THE UP AND OK SYMBOLS WERE FOUND TO BE WORN. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE PUMP BOOTED TO THE VERIFY SCREEN WITH DIM PINK CONTRAST. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303387 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 41 YR