FDA Adverse Event Malfunction Summary report: N

ENDEAVOR RESOLUTE RX

MDR report key: 3200841 · Received July 2, 2013

Report

Report Number
9612164-2013-00682
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
May 29, 2013
Report Date
June 24, 2013
Manufacturer
MEDTRONIC IRELAND
Product Code
NIQ
PMA / PMN Number
P110013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RESULTS: INHERENT RISK OF PROCEDURE ¿ (FAILURE TO DELIVER AND STENT DEFORMATION). PATIENT¿S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (LESION MORPHOLOGY) ¿ CTO LESION. (B)(4).

Description of Event or Problem · 1

PHYSICIAN WAS ATTEMPTING TO DEPLOY ONE ENDEAVOR RESOLUTE DRUG-ELUTING STENT TO THE RCA (CTO LESION) BUT THE DEVICE COULD NOT CROSS. LESION HAD BEEN PRE-DILATED. NO PATIENT COMPLICATION WAS REPORTED. EVALUATION SUMMARY: THE STENT WAS POSITIONED BETWEEN THE MARKER BANDS AS PER SPECIFICATIONS. THE 10TH AND 11TH PROXIMAL SEGMENTS WERE RAISED AND DEFORMED. A NUMBER OF OTHER SEGMENTS WERE SLIGHTLY RAISED AND MISALIGNED. THE DISTAL TIP WAS DAMAGED. PLEASE NOTE THAT THIS DEVICE (ENDEAVOR RESOLUTE) IS DISTRIBUTED OUTSIDE THE UNITED STATES; HOWEVER, IT IS SIMILAR TO THE UNITED STATES DISTRIBUTED PRODUCT (RESOLUTE INTEGRITY).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303341 ENDEAVOR RESOLUTE RX STENT, CORONARY, DRUG-ELUTING NIQ MEDTRONIC IRELAND 0006585080

Patients

Seq Age Sex Outcome Treatment
1 00069 YR