ENDEAVOR RESOLUTE RX
Report
- Report Number
- 9612164-2013-00682
- Event Type
- Malfunction
- Date Received
- July 2, 2013
- Date of Event
- May 29, 2013
- Report Date
- June 24, 2013
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- NIQ
- PMA / PMN Number
- P110013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
RESULTS: INHERENT RISK OF PROCEDURE ¿ (FAILURE TO DELIVER AND STENT DEFORMATION). PATIENT¿S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (LESION MORPHOLOGY) ¿ CTO LESION. (B)(4).
PHYSICIAN WAS ATTEMPTING TO DEPLOY ONE ENDEAVOR RESOLUTE DRUG-ELUTING STENT TO THE RCA (CTO LESION) BUT THE DEVICE COULD NOT CROSS. LESION HAD BEEN PRE-DILATED. NO PATIENT COMPLICATION WAS REPORTED. EVALUATION SUMMARY: THE STENT WAS POSITIONED BETWEEN THE MARKER BANDS AS PER SPECIFICATIONS. THE 10TH AND 11TH PROXIMAL SEGMENTS WERE RAISED AND DEFORMED. A NUMBER OF OTHER SEGMENTS WERE SLIGHTLY RAISED AND MISALIGNED. THE DISTAL TIP WAS DAMAGED. PLEASE NOTE THAT THIS DEVICE (ENDEAVOR RESOLUTE) IS DISTRIBUTED OUTSIDE THE UNITED STATES; HOWEVER, IT IS SIMILAR TO THE UNITED STATES DISTRIBUTED PRODUCT (RESOLUTE INTEGRITY).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 303341 | ENDEAVOR RESOLUTE RX | STENT, CORONARY, DRUG-ELUTING | NIQ | MEDTRONIC IRELAND | 0006585080 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00069 YR |