FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 3200747 · Received July 1, 2013

Report

Report Number
1416980-2013-16973
Event Type
Malfunction
Date Received
July 1, 2013
Date of Event
June 7, 2013
Report Date
June 7, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RECEIVED ONE UNUSED MINI-CAP TRANSFER SET FOR EVALUATION. VISUAL INSPECTION, LEAK TESTING AND CLEAR PASSAGE TESTS WERE PERFORMED WITH NO ISSUES NOTED. CLAMP FUNCTION TEST PERFORMED WITH NO ISSUES NOTED. PRESSURE TESTED THE SAMPLE WITH ACCEPTABLE RESULTS. FUNCTIONALLY HAND TIGHTENED A LAB TITANIUM ADAPTER TO SET WITH DIFFICULTY NOTED. DIMENSIONAL INSPECTION OF THE RETURNED TRANSFER SET IDENTIFIED THAT THE INSIDE DIAMETER OF THE CATHETER ADAPTER SHROUD WAS BELOW NOMINAL. THE SAMPLE WAS CONFIRMED FOR THE REPORTED CONDITION. IMPROVEMENTS WERE MADE TO THE MOLD AND MOLDING DEPARTMENT PROCEDURES. A REVIEW OF ALL BATCH RECORD DOCUMENTS LOT NUMBER H12J01053 WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF RETURNED COMPANION SAMPLES, A MINICAP TRANSFER SET WAS FOUND TO NOT MEET DIMENSIONAL SPECIFICATIONS. THIS SAMPLE HAD BEEN RETURNED UNOPENED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
297761 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H12J01053

Patients

Seq Age Sex Outcome Treatment
1