UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM
Report
- Report Number
- 2122870-2013-00584
- Event Type
- Malfunction
- Date Received
- July 1, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 3, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- DHA
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) PERFORMED A FAILING DILUTION TEST AND DETERMINED POSSIBLY FAULTY PRECISION PUMP. THE FSE REBUILT THE PRECISION PUMP AND VERIFIED PIPETTE ALIGNMENTS, ULTRASONIC SETTINGS, AND MIXER SPEEDS. THE FSE REPEATED THE DILUTION TEST WITH PASSING RESULTS AND PERFORMED TESTING OF THE ELEVATED PATIENT SAMPLE AND GENERATED DIL-BHCG RESULTS THAT MATCHED THE DIL-BHCG RESULTS FROM THE ALTERNATE SECOND ACCESS INSTRUMENT AND RESULTS FROM MANUALLY DILUTED SAMPLES. THE FSE PERFORMED PASSING QUALITY CONTROL (QC) TO VERIFY INSTRUMENT PERFORMANCE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION.
THE CUSTOMER REPORTED SUPPRESSED DIL- BETA HUMAN CHORIONIC GONADOTROPHIN (BHCG) RESULTS (<1,000M IU/ML) FOR ONE PATIENT SAMPLE AND SEVERAL SURVEY SAMPLE RESULTS INVOLVING THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM. REPEAT DIL-BHCG TESTING OF THE SAMPLES, ON AN ALTERNATE ACCESS INSTRUMENT, PRODUCED EXPECTED RESULTS THAT WERE GREATER THAN 1,000 MIU/ML AND CORRELATED WITH THE ORIGINAL BHCG RESULTS (>1,000 MIU/ML). THE CUSTOMER STATED THE ERRONEOUSLY SUPPRESSED DIL-BHCG PATIENT RESULT OF <1,000 MIU/ML WAS NOT RELEASED FROM THE LABORATORY. THERE WAS NO PATIENT CONSEQUENCE OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. QUALITY CONTROL (QC) WAS WITHIN SPECIFICATION AT THE TIME OF THE EVENT. SYSTEM CHECK, PERFORMED ON THE DAY OF THE EVENT, PASSED WITHIN SPECIFICATIONS. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 297759 | UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | DHA | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |