PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3200581 · Received July 1, 2013

Report

Report Number
3004209178-2013-95289
Event Type
Injury
Date Received
July 1, 2013
Date of Event
June 10, 2013
Report Date
June 12, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 469MG/DL. THE CALLER STATED THAT HER INFUSION SET FELL OUT WHEN SHE WAS SLEEPING. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. THE TIME, DATE, BASAL RATES, AND BOLUS WIZARD SETTINGS WERE CORRECT. ASSISTED THE CUSTOMER TO RUN A MANUAL PRIME AND THE INSULIN DID EXIT. ADVISED THE CUSTOMER TO CALL BACK WHEN THE TUBING CLAMP ARRIVES TO CONTINUE TESTING. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299753 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 21 YR Hospitalization