PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3200557 · Received July 1, 2013

Report

Report Number
3004209178-2013-95264
Event Type
Malfunction
Date Received
July 1, 2013
Date of Event
June 11, 2013
Report Date
June 25, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH OPERATING CURRENTS WITHIN SPECIFICATIONS. THE DEVICE RECEIVED WITH AN INTERMITTENT BUTTONS RESPONSE DUE TO CORRODED KEYPAD TRACES. UNABLE TO CONFIRM THE BUTTON ERROR ALARMS. THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE/FLUSH DRIVE SUPPORT DISK. THE UNIT WAS RECEIVED WITH SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE BLOOD GLUCOSE READING WAS 142MG/DL. THE CUSTOMER STATED THAT THE BATTERY WAS REPLACED, BUT THE ANOMALY CONTINUED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300108 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 50 YR