FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3200557
·
Received July 1, 2013
Report
- Report Number
- 3004209178-2013-95264
- Event Type
- Malfunction
- Date Received
- July 1, 2013
- Date of Event
- June 11, 2013
- Report Date
- June 25, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH OPERATING CURRENTS WITHIN SPECIFICATIONS. THE DEVICE RECEIVED WITH AN INTERMITTENT BUTTONS RESPONSE DUE TO CORRODED KEYPAD TRACES. UNABLE TO CONFIRM THE BUTTON ERROR ALARMS. THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE/FLUSH DRIVE SUPPORT DISK. THE UNIT WAS RECEIVED WITH SCRATCHED DISPLAY WINDOW.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE BLOOD GLUCOSE READING WAS 142MG/DL. THE CUSTOMER STATED THAT THE BATTERY WAS REPLACED, BUT THE ANOMALY CONTINUED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300108 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |