FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3200353 · Received July 1, 2013

Report

Report Number
1416980-2013-16962
Event Type
Malfunction
Date Received
July 1, 2013
Date of Event
June 6, 2013
Report Date
June 6, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE BY A FIELD SERVICE TECHNICIAN. REVIEW OF THE EVENT LOG IDENTIFIED THE REPORTED F-38 ALARM. THE CAUSE OF THE ALARM WAS DETERMINED TO BE THAT THE RIGHT FORCE SENSING RESISTOR (FSR) WAS DAMAGED. IN ORDER TO ADDRESS THIS CONDITION THE FSR WAS REPLACED. THE DEVICE WAS RESTORED TO GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, A FOLLOW-UP WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED AN F-38 ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298678 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1