HOMECHOICE PRO
MDR report key: 3200336
·
Received July 1, 2013
Report
- Report Number
- 1416980-2013-16961
- Event Type
- Malfunction
- Date Received
- July 1, 2013
- Date of Event
- November 14, 2012
- Report Date
- June 6, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
A HIGH DRAIN ERROR 119 ALARM (DURING NIGHT DRAIN #19) WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THIS ALARM INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE. THE ALARM OCCURRED ON (B)(6) 2012 AT 20:00:25. NO ADDITIONAL INFORMATION IS AVAILABLE.