FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3200312 · Received July 1, 2013

Report

Report Number
1416980-2013-16957
Event Type
Malfunction
Date Received
July 1, 2013
Date of Event
June 6, 2013
Report Date
June 6, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED, A DEVICE ANALYSIS COULD NOT BE COMPLETED. PROPER USER INSTRUCTIONS ARE ADDRESSED IN "THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE" WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE HAS INSTRUCTIONS ON THE STEPS TO PROPERLY DISCONNECT FROM CYCLER DURING AN EMERGENCY AND EXPLAINS HOW TO RECONNECT TO THERAPY AFTER PROPERLY DISCONNECTING. A LABELING REVIEW OF THE PRODUCT FAMILY WILL BE PERFORMED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM OCCURRED ON THE HOMECHOICE (HC) MACHINE. THIS OCCURRED DURING DRAIN ONE, WHILE THE PATIENT WAS CONNECTED. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM. THE HOME PATIENT (HP) STATED THAT THEY HAD DISCONNECTED AND RECONNECTED. HOWEVER, THE PROPER DISCONNECTION PROCEDURE WAS NOT USED. THE TSR ASSISTED WITH THE TROUBLESHOOTING IN ORDER TO CLEAR THE ALARM. THE TSR REVIEWED THE PROPER PROCEDURE WITH THE HP. THE HP WAS ADVISED TO DISCARD THE SUPPLIES IN ORDER TO START OVER WITH NEW SUPPLIES OR FINISH THERAPY WITH MANUAL BAGS. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299673 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE