FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3200308 · Received July 1, 2013

Report

Report Number
2032227-2013-02680
Event Type
Malfunction
Date Received
July 1, 2013
Date of Event
June 27, 2013
Report Date
June 27, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE LEVELS AND NO DELIVERY ALARMS. THE CUSTOMER CHANGED THE INFUSION SET TWICE, BUT STILL GOT THE ALARM. IT WAS ALSO STATED THAT THE DRIVE SUPPORT CAP WAS FLUSH. TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP PASSED THE DISPLACEMENT AND SELF TESTS. DURING THE PRIME TEST, INSULIN EXITED, THEN STOPPED ABOUT HALF WAY THROUGH THE TEST. NOTHING FURTHER WAS REPORTED.