FDA Adverse Event
Malfunction
Summary report: N
11CM ANGLE ATTACHMENT
MDR report key: 3200025
·
Received July 1, 2013
Report
- Report Number
- 1045834-2013-02624
- Event Type
- Malfunction
- Date Received
- July 1, 2013
- Date of Event
- May 30, 2013
- Report Date
- June 7, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBE
- PMA / PMN Number
- K002903
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED. EVIDENCE INDICATES THIS WAS DUE TO USAGE WEAR OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A SPINE SURGERY THE ATTACHMENT DEVICE "OVERHEATED". THE PLANNED SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT DELAY AS A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS ALLEGED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 299914 | 11CM ANGLE ATTACHMENT | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)-ATTACHMENT | HBE | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |