FDA Adverse Event Malfunction Summary report: N

SYNCHROMED II

MDR report key: 3200013 · Received July 1, 2013

Report

Report Number
3007566237-2013-01099
Event Type
Malfunction
Date Received
July 1, 2013
Date of Event
February 16, 2005
Report Date
February 16, 2005
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
LKK
PMA / PMN Number
P860004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

A PROBLEM WAS REPORTED RELATED TO PUMP. A MOTOR STALL WAS REPORTED AFTER ATTEMPTING TO READ THE SYNCHROMED II PUMP WITH A MAGNET. MOTOR STALL CONFIRMED IN LOGS AND RECOVERED. COMPANY REPRESENTATIVE REPORTED HCP INADVERTENTLY USED MAGNET TO INTERROGATE THE PUMP CAUSING MOTOR STALL. PUMP WAS RUNNING AT APPROXIMATELY 0.2ML/DAY. NO PATIENT SYMPTOMS OR OUTCOME WERE REPORTED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A REPORT WILL BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300340 SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MEDTRONIC NEUROMODULATION 8637

Patients

Seq Age Sex Outcome Treatment
1