ENDO ILS 21MM, CURVED
Report
- Report Number
- 3005075853-2013-03374
- Event Type
- Malfunction
- Date Received
- July 1, 2013
- Date of Event
- June 18, 2013
- Report Date
- June 18, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K940967
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION WAS NOT PROVIDED BY CONTACT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED WITH THE ANVIL MISSING, OTHERWISE IN GOOD VISUAL CONDITION. AS THE ORIGINAL ANVIL WAS NOT RECEIVED, FURTHER INVESTIGATION WITH THE ORIGINAL ANVIL COULD NOT BE PERFORMED. THE BREAKAWAY WASHER WAS NOT PRESENT AND THERE WERE NO STAPLES PRESENT. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY WITH A TEST ANVIL; IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A GASTRIC BYPASS PROCEDURE, GJ ANASTOMOSES DEVICE FIRED BUT TISSUE WASN'T FULLY CUT. THE FIRST ASSIST FIRED THE DEVICE. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. THERE WERE NO PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 298657 | ENDO ILS 21MM, CURVED | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |