KNIFE
Report
- Report Number
- 2523835-2013-00089
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Date of Event
- April 18, 2013
- Report Date
- May 17, 2013
- Manufacturer
- ALCON PRECISION DEVICE - SINKING SPRING
- Product Code
- HNN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ONE OPENED SAMPLE WAS RETURNED. EVALUATION OF THE SAMPLE SHOWED THE FOLLOWING RESULTS: THE SAMPLE WAS EXAMINED USING (B)(4) MAGNIFICATION AND FOUND TO HAVE A DAMAGED TIP AND CUTTING EDGES. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THIS LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO THE MANUFACTURER'S ACCEPTANCE CRITERIA. THE EXACT ROOT CAUSE FOR THE DAMAGED KNIFE IS UNK; HOW OR WHEN THE BLADE BECAME DAMAGED CANNOT BE DETERMINED. THE DAMAGE TO THE RETURNED SAMPLE IS CONSISTENT WITH DAMAGE THAT CAN OCCUR WHEN THE BLADE CONTACTS ANOTHER SURFACE PRIOR TO USE. SOME POTENTIAL CAUSES COULD BE REUSE, IMPROPER HANDLING, OR CONTACT WITH ANOTHER INSTRUMENT ON THE INSTRUMENT TRAY DURING PROCEDURE SETUP. ALL KNIVES ARE 100 PERCENT INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF (B)(4) MAGNIFICATION DURING MANUFACTURING. ANY DEFECTS, SUCH AS THE DAMAGED TIP AND CUTTING EDGES EXHIBITED ON THE RETURNED OPENED SAMPLE, ARE REMOVED FROM THE LOT AND SCRAPPED. PENETRATION TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. (B)(4).
A CUSTOMER REPORTED THAT A KNIFE WAS DULL AND WOULD NOT CUT DURING A PROCEDURE. THE CASE WAS COMPLETED WITH NO HARM TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291573 | KNIFE | MANUAL OPHTHALMIC SURGICAL INSTRUMENT | HNN | ALCON PRECISION DEVICE - SINKING SPRING | 8065982465 | 904894M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CLEARCUT HP2 DB SLIT 2.4MM ANGLED KNIFE |